XT V 160 X-ray System XTM0011-F2 75
3. In the Properties window, click to select the X-ray properties.
4. Click the Alignment tab.
5. Ensure that the Auto filament control check box is selected.
6. Click Calibrate.
A message appears asking whether you have just performed filament change.
7. Click No to display the Filament Set-Up window.
Ensure that a full calibration has been run previously before you click No.
8. If you do not want to auto-condition the source, clear the Auto-condition check box.
9. Ensure there is no sample in view and click Start.
Inspect-X auto-conditions the source (optional), calibrates the filament demand and the
sets up the X- and Y-shifts.