4-8
I4-1
OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
1. All gaps in the sensor module must not be blocked.
2. Please pull out and plug in the sensor module correctly.
3. Sensor Module is an important part which supply Temperature Control to
Incubator, and must operate carefully.
H. CHECK FAN MOTOR ALARM
Block the inlet port and outlet port on the left and right side of main deck, after several
minutes, the device gives alarms, the “Alarm” light will flash with continuous audible sound, and
the Message indicator will show “Airflow Error”.
Turn off the power supply of controller, and pull off the power cord, and take out the
controller.Reconnect the power cord, and hold the fan slightly, and turn on the switch of controller,
and the fan motor alarm occurs, the “Alarm” light will flash with continuous audible sound, and
the Message indicator will show “Motor Error”.
WARNING: 1. If the temperature of heater inside controller is too high, you can
remove the conteoller or touch the heater after the unit is switched
off at least 45min to avoid the burn.
2、starting after taking out the controller, please pay attention to the
power plug of oxygen concentration and humidity control system
to avoid electrical shock.
I. CHECK OVER-TEMPERATURE ALARM
In Air Mode, enter into the working state, press Up, Down and Silence/Reset key, there is
no indication on the Set Temperature indicator, and the heat power indicator is on in full, and the
controller has entered into the overtemp checkout state, and when the incubator temperature is
less than 38.0℃, the over-temperature alarm should active. The “Alarm” light will flash with
continuous audible sound, and the Message indicator will show “Over Temp”. In the state of >
37℃ Override Mode, enter into the working state, press Up, Down and Silence/Reset key,
there is no indication on the Set Temperature indicator, and the heat power indicator is on in full,
and the controller has entered into the overtemp checkout state, and when the incubator
temperature is less than 40.0℃, the over-temperature alarm should active. The “Alarm” light will
flash with continuous audible sound, and the Message indicator will show “Over Temp”.
In Baby Mode, press Up, Down and Silence/Reset key, there is no indication on the Set
Temperature indicator, and the heat power indicator is on in full, and the controller has entered
into the overtemp checkout state, and when the incubator temperature is less than 40.0℃, the
over-temperature alarm should active. The “Alarm” light will flash with continuous audible sound,
and the Message indicator will show “Over Temp”.