4-9
I4-1
OPERATOR’S MANUAL FOR INFANT INCUBATOR
EDITION/REVISION B/1
J.CHECK DEVIATION ALARM
In Air Mode, close all door and panelss, set the temperature at 32 ºC. Enter
TEMPERATURE ALARM CHECKOUT STATE(refer to descriptions of terms and symbols
concerned), fan into the hot air inside the hood. When the air temperature indicates 35.1ºC, the
high deviation alarm will active. The “Alarm” light will flash with continuous audible sound, and
the Message indicator will show “High Temp”; set the temperature at 35℃, after the device
enters into the temperature alarming checkout state, open the front access panel, when the air
temperature indicates 31.9 ºC, the low deviation alarm should active. The “Alarm” light will flash
with continuous audible sound, and the Message indicator will show “Low Temp”.
NOTE: If the system can not enter into the TEMPERATURE ALARM CHECKOUT STATE or
the air Temperature does vary wihtin ±3℃ than the setting temperature, the
deviation alarm can not occur.
In Baby Mode, set the temperature at 35 ºC. Enter TEMPERATURE ALARM CHECKOUT
STATE, put the skin sensor into the water cup at 37 ºC. When the skin temperature indicates 36.1
ºC, the high deviation alarm will active. The “Alarm” light will flash with continuous audible sound,
and the Message indicator will show “High Temp”; Set the temperature at 35℃, after the device
enters into the temperature alarming checkout state, put the skin sensor into the water cup at
temperature 33℃, when theskin temperature indicates to 33.9ºC, the low deviation alarm should
active. The “Alarm” light will flash with continuous audible sound, and the Message indicator will
show “Low Temp”.
NOTE: If the system can not enter into TEMPERATURE ALARM CHECKOUT STATE or the
skin temperature does vary within ±1℃ than the setting temperature, the
deviation alarm can not occur.
K. CHECK THE WATER LACKING ALARM
When humidifying system is on working, pour out the water in the water tank. After a while,
the incubator will alarm. The “Alarm” light will flash with continuous audible sound, and the
Message indicator will show “Low Water”.
L. CHECK THE WATER TANK IN INCORRECT LACATION ALARM
When the incubator is in the working state, pulling out the water tank, the incubator will alarm.
The “Alarm” light will flash with continuous audible sound, and the Message indicator will show
“Tank Position”.
M. CHECK THE TIMER
When the controller is normally working, the Message shall diaplays the current realtime. If
the displayed time is not correct, please reset the time. The seting method is as following: