5-5
I4-1
OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/1
F. Remove Air Intake Tube Connector
See Figure 5.6, Move the Air Intake Tube out in direction of arrow 2 after disconnecting Air
Intake Tube from Air Tube Connector in direction of arrow 1
FIGURE 5.6
G. Remove Reservoir
Pull the Reservoir out slowly till it is locked as the figure 5.7 directs, then press it down
slightly, it can be pulled out.
Affusion meatus
Reservoir
FIGURE 5.7
CAUTION:1. Because the temperature of the water tank is high after the power of the
device is on as well as the heating groups are tighten then, so please take out
the water tank after it is totally cooling down to avoid damage to the heating
group.
2. The humidifying system is in the bottom of the reservoir. Don’t put hands
into the reservoir to avoid being burned.
H. Remove the Access door gasket and Access door cuff, and Iris port sleeve from hood.
I. Remove the soft tubing from hood.