OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/1
5.2.2 CLEANING PROCEDURE
CAUTION: Some chemical cleaning agents may be conductive and/or leave a residue
which may permit a build-up of dust or dirt which may be conductive. Do not
permit cleaning agents to contact electrical components. Do not spray
cleaning solutions onto any of those surfaces.
A. Clean the skin temperature sensor and sensor modular
Use a national registered neutral disinfectant-detergent to thoroughly clean all surfaces; then
dry with a clean cloth or paper towel.
CAUTION: Do not steep Sensor into disinfectant-detergent and avoid the
disinfectant-detergent leak into the sensor modular.
B. Clean the Soft tubing, Gaskets, Iris Port Sleeves
Allow parts to soak as national registered neutral disinfectant-detergent, then remove them
and dry completely with a clean cloth or paper tower.
C. Clean the Hood and inner wall
Use a national registered neutral disinfectant-detergent to thoroughly clean all surfaces,
including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or
paper towel.
NOTE: Alcohol can cause crazing of the clear Acrylic panel. Do not use alcohol, acetone,
or any organic solvents for cleaning. Do not expose the panel assembly to direct
ultraviolet radiation.
D. Clean the controller
Use national registered neutral disinfectant-detergent to clean the all surfaces thoroughly.
Then dry it with clean cloth.
NOTE: Pay particular attention to avoid liquid seeping into the Controller Shell.
Please clean the heater after it is totally cooled down.
WARNNING:Not clean the heater and fan will result in accumulate of fibre dust, baffle the
air flow. That will infulence the temperature control and result in high CO
2
concentration.
E. Clean the bassinet, Main Deck, shelf and tray
Use a national registered neutral disinfectant-detergent to thoroughly clean all surfaces,
including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or
paper towel.
CAUTION: Do not lubricate the Mattress Tilt Mechanism with oil or other potentially
F. Clean mattress cover
Use a national registered neutral disinfectant-detergent to thoroughly clean it and sun it after
washing it with clean water.
G. Clean Air filter
Use a national registered neutral disinfectant-detergent to thoroughly clean all surfaces,
including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or
paper towel. If the air filter is dirt or is used for more than 2 months, please replace it.