5-7
I4-1
OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/3
H. Clean Water tank and Humidity chamber fill spout
Open the two cover boards on the humidifying chamber and pull up the water proof as
figure5.8 indicates. Use the national registered neutral disinfectant-detergent to clean the all
surfaces and cupped place thoroughly. Clean it again with clean water and then dry it with clean
cloth or paper.
Please pay special attention that clean the inner and outter surface of the probe with the
around the detect point can control the microorganism breed in the water tank and ensure the
accuracy of the water level detecting.
When the scale appears visibly on the radiating cover inside the water tank, it is
recommended to soak it with table vinegar or the scale cleaning detergents which is used
exclusively for casting water dispenser. The soaking time is about 15min to 20 min. (The longer
the radiating cover be soaked, the better) .In the process of soaking, make sure that the cover is
immersed entirely. After soaking, scrub with bottle brush and rinse the cover till the scale is
cleaned completely.
Cover board
Reek outlet Detecting probe
Radiating cover
Water proof
Water inlet hole on the bottom
CAUTION:1. Please clean the water tank after the radiating cover inside the water tank
totally cooleddown.
2. Make sure the water inlet hole on the bottom and the reek outlet hole of tank
throughable after cleaning.
3. If the water tank is uncleaned, especially there're dusts near the probe, there
will be a possibility that water-lacking alarm failure.
4. Please use soft cloth to wipe the surface of the probe when it is covered by
nobel metal. Do not use the hard material to scrape the probe so as not to
damage it.
5. When the scale appears on the radiating cover visibly, you must clean it in
time, or it will easily breed ground for bacterium and affect the humidification.
FIGURE 5.8
I. Clean NEONATE BILIRUBIN OHOTOTHERAPY EQUIPMENT
See the user’s manual for XHZ-90 and XHZ-90L.
J. Clean surface of device
Use a national registered neutral disinfectant-detergent to thoroughly clean all surfaces,
including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or
paper towel.
5.2.3 REASSEMBLY AFTER CLEANING
NOTE: Before reassembling parts into Incubator, check carefully if there is any crack or
damage. Some disinfectant-detergent may cause crazing of plastic parts in
Incubator.
A. Install Humidity chamber fill spout in the reverse steps as they are moved out.