Components
No. Key Name Display Function
1
Power Key -
Press once to turn ON the power. When the power is ON,
hold it down for more than two seconds to turn OFF the
power
2
Memory Key
G
In General Application Mode:
Motor speed can be adjusted between 100 and
40,000 min
-1
.
In Rotary Endo Mode:
Motor speed can be adjusted between 100 and
6,000 min
-1
.
*Display value changes interlocked with the gear ratio.
3
Torque Key
(NLZ E)
B
When gear ratio is set by each time pressing this key you
can adjust the torque.
Rotary Endo Mode
Gear Ratio Torque Range (N·cm) Increment (N·cm)
1:1 0.3 - 3.0 0.1
4:1 0.8 - 6.0 0.4
6:1 1.0 - 6.0 1.0 - 1.2:0.2
1.2 - 6.0:0.6
-
Change the setting as described in “3-8 Changing the
settings for various functions”
Setting Key
(NLZ)
-
Change the setting as described in “3-8 Changing the
settings for various functions”
4
Speed Key
A
In General Application Mode:
Motor speed can be adjusted between 100 and
40,000 min
-1
.
In Rotary Endo Mode:
Motor speed can be adjusted between 100 and
6,000 min
-1
.
*Display value changes interlocked with the gear ratio.
5
Check Key
C
Check the status of the handpiece as described in “4-4
Contra-Check Function”
6
Light Key
D
Set light intensity (Strong·Moderate·Weak·OFF)
By each time pressing this key you can change the
volume from OFF -> intensity 1 -> intensity 2 -> intensity
3 and then back to OFF again
* The light can be turned on for 5 seconds by pressing the
key while the motor is stopped.
10