Post-use Maintenance
5
CAUTION
ŘFollow local rules, regulations, and guidelines regarding the reprocessing of devices.
ŘPerform the following cleaning and disinfection quickly after treatment (within 1 hour) to remove any residue.
ŘDo not perform steam sterilization the control unit, motor cord and AC adapter, AC power cord.
ŘDo not lubricate the motor. It may cause overheating and product failure.
ŘDo not use the following fluids to wipe, immerse or clean the product; strong/super acid water, strong acid/
alkaline chemicals, chlorine containing solutions, solvents such as benzine or thinner. It may cause color change
of resin part, or corrosion of the metal part of the NLZ Motor System.
ŘSteam sterilization is recommended for the product. The validity of other sterilization methods (such as plasma
sterilization or EOG sterilization) is not confirmed.
ŘFor details on maintenance of the handpiece, check the Operation Manual of the handpiece.
ŘDo not immerse the product in disinfectant or perform cleaning with an ultrasonic washer. Doing so may cause
product failure.
After each patient, maintain the product as follows.
5-1
Cleaning at point-of use (Motor)
1
Always wear protective gloves, a mask, and
protective goggles for safety purposes and to
minimize the risk of infection.
2
Turn off power by pressing the Power Key
2
3
Remove the bur/file.