Operation Procedure
23
4-1
General Application Mode
1
2
5
4
Select the proper speed required for the application
3
6
Set the light intensity
Gear Ratio Rotation Speed (min
-1
) Setting range (min
-1
)
1:5 500 – 200,000 20 – 100:10 Increment
1:1 100 – 40,000
100 – 1,000:100 Increment
1,000 – 5,000:500 Increment
4:1 30 – 10,000
5,000 – 50,000:1,000 Increment
6:1 20 – 6,600 50,000 – 200,000:5,000 Increment
Select the gear ratio
according to the handpiece
1:5, 1:1, 4:1, 6:1
Select the program # - 6
programs can be selected
Select the General
Application Mode
(NLZ E only)
Select the rotation
direction, FWD / REV
Strong
Moderate
Weak
OFF
NOTICE
ŘSince the NLZ Endo contra angle handpiece is only for endodontic treatment, it cannot be used for General
Application Mode.
ŘThe rotation speed of the bur, mounted to the handpiece, depends on the gear ratio of the handpiece.
ŘThe actual rotation speed is displayed during operation. The Max set speed is displayed when stopped.
ŘThe max set speed can be adjusted even during rotation by pressing the Speed Keys on the control panel.
7
Adjust the water spray of the handpiece at the dental unit
8
Contra-Check
Perform Contra-Check before treatment. (Refer to “4-4 Contra-Check Function”)