6
1-4
Classification of Equipment
ŘType of protection against electric shock:
- Class ll equipment
ŘDegree of protection against electric shock:
- Type B applied part:
(Applied parts: Motor, Handpiece)
ŘMethod of sterilization or disinfection recommended by the manufacturer:
- Refer to ”5-3 Packaging, Sterilizing, Drying and Storage (Motor)”
ŘDegree of protection against ingress of water as detailed in the current edition of IEC 60529:
- Control Unit: IPX0 (Not protected)
ŘDegree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen
or nitrous oxide:
- Equipment NOT suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen
or nitrous oxide.
ŘMode of operation:
- Intermittent operation ( ON:3 min, OFF:10min.)