16
Preparations for Use
3-7
Connecting / disconnecting the motor and the handpiece (option)
<Connecting>
1
Insert the E-type handpiece into the motor insert,
and turn until it clicks as the positioning pin on
the handpiece falls into positioning hole on the
motor.
2
Confirm that the handpiece is securely connected
to the motor.
MotorHandpiece (option)
<Disconnecting>
Simply pull out the handpiece from the motor.
CAUTION
ŘEnsure that the motor has completely stopped rotating before handling. Connecting or disconnecting the
handpiece while the motor is rotating may result in injuries or damage the handpiece.
ŘBe sure to adjust the rotation speed of the motor within the allowable rotation speed of the handpiece (option).
ŘLubricated handpiece should stand and allow oil to drain prior to steam sterilization cycle. Attach to the motor
after the excess oil has been completely drained. If the oil enters the motor, it may cause malfunction of the
motor.
ŘAfter lubrication, keep the handpiece standing apart from the motor. Connect the handpiece to the motor when
using.
ŘDo not allow water to enter the motor. It may cause malfunction of the motor.
ŘBefore each use, operate the motor outside the patient's oral cavity. If any abnormality, such as vibration, noise
or overheating occurs, stop using the product immediately and contact your Authorized NSK Dealer. (Refer to
"3-9 Check before treatment")