Operation Procedure
26
<Auto Reverse Function>
By each time pressing the Auto Rev. Key you can switch the mode between AUTO STOP -> AUTO
REVERSE STOP -> AUTO REVERSE FORWARD and then back to AUTO STOP again
AUTO STOP AUTO REVERSE STOP AUTO REVERSE FORWARD
Explaining the above 3 Modes:
AUTO STOP
The handpiece starts in forward rotation. When a load reaches the set torque limit value, the motor rotation
stops with a beep sound. If the load continues, the set speed on the LCD Display starts blinking until the
foot pedal is released.
Load lower than
the set torque limit value
Forward rotation Stop
When load reaches
the set torque limit value
AUTO REVERSE STOP
The handpiece starts in forward rotation. When a load higher than the torque limit is applied, the file will
rotate in reverse with a beep sound. When the load is removed, the motor rotation stops and the set speed
on the LCD Display starts blinking until the foot pedal is released. If you want the motor to rotate again, re-
press the foot pedal.
Load lower than
the set torque limit value
Reverse rotation when load continues,
stop when load is removed
Forward rotation Reverse rotation Stop
Load higher than
the set torque limit value