EasyManua.ls Logo

OCULUS Myopia Master Advanced - Page 48

OCULUS Myopia Master Advanced
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 Measuring Procedure
42 / 92 Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
Refraction Values (2)
The sphere, cylinder, axis position and quality values are displayed in this
field.
SEQ: Spherical equivalent
Q-value:
If the field has a white background (9-8) - the measuring results are good.
If the field has a yellow background (7-6) - the measuring results are
critical; repeat the measurement, if necessary.
If the field has a red background (
6) - repeat the measurement.
AXL values (3)
The axial length values for one or both eyes are shown.
The corresponding signal to noise ration (SNR) is listed.
Keratometer Values (6)
K1/K2: Horizontal/vertical radius of curvature in the center
blue: flat meridian
red: steep meridian
Pupil: Size of pupil
Astig: Astigmatism of cornea in the center
WTW: (white-to-white) Cornea diameter or iris diameter.
Q-value:
If the field has a white background (9-8) - the measuring results are
good.
If the field has a yellow background (7-6) - the measuring results are
critical; repeat the measurement, if necessary.
If the field has a red background (
6) - repeat the measurement.

Table of Contents

Related product manuals