EasyManua.ls Logo

ODA OPS Series - Measure Commands

Default Icon
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FLOWer?
Command that checks power supply's current CV(Constant Voltage, CC(Constant Current).
Return value "CV" - Constant Voltage Condition
"CC" - Constant Current Condition
ex) FLOW?
return value "CV"
POLarity {P|N}
Command is only for Polarity Power Supply, changes polarity of output terminal.
> P Change to straight polarity output.
> N Change to reverse polarity output.
ex) POL P
Change to straight output
POLarity?
Command that checks power supply's polarity condition. Only used for Polarity Power Supply.
Return value "P" - Straight polarity output condition.
"N" - Reverse polarity output condition.
ex) POL?
return value "P"
6-5. Measure Command
Command that measures power supply's read back voltage & current. Does not require
DVM(Digital Volt Meter) & Ammeter.
MEASure:VOLTage[:DC]?
Command that measure power supply
'
s output voltage
Command
that
measure
power
supply s
output
voltage
.
Return value "voltage"
ex) meas:volt
?
return value "11.0000"
MEASure:CURRent[:DC]?
Command that measure power supply's output current.
Return value "current"
ex) meas
:
curr?
return value
"
1.0000
"
ex)
meas:curr?
return
value
1
.
0000
- 69 -