7-3. Remote Calibration Error
Read reference "4-6. Calibration(for GPIB) using REMOTE INTERFACE"
-20, "Ignored min run under volt"
This error occurs when Max or value is activated while Min voltage value is not operated.
Sequence : Min → VALUE → MAX → VALUE
-21, "Ignored min save under volt"
Operated MAX Value without voltage min value.
Sequence : Min → VALUE → MAX → VALUE
-22, "Invalid min value use under volt"
This error occurs when you have send value without activating max value after operating
minumum value.
Sequence : Min → VALUE → MAX → VALUE
-23, "En route to cal the curr"
This error occurs when you have send voltage calibration while current calibration.
-
"
"
,
This error occurs when Minimum value is out of range.
Read reference "4-5. CALIBRATE Setting Range"
-25, "Under volt max parameter"
This error occurs when voltage max value range is out of lowest limit.
Read reference "4-5. CALIBRATE Setting Range"
-26, "Over volt max parameter"
This error occurs when voltage max value range is out of highest limit.
Read reference "4-5. CALIBRATE Setting Range"
-27, "Ignored min run under curr"
This error occurs when max value is operated when current min value is not operated.
Se
uence : Min → VALUE → MAX → VALUE
-28, "Ignored min save under curr"
Operated MAX value before operating current min value.
Sequence : Min → VALUE → MAX → VALUE
-29, "Invalid min value use under curr"
This error occurs when you have sent value again without o
erating maximum after minimu
This error occurs when you have sent value again without operating maximum after minimum
Sequence : Min → VALUE → MAX → VALUE
- 76 -