KEYLock[:STATe] {OFF|ON|0|1}
Command that controls front panel's key & encorder switch.
> ON OR 1 Stop Operating
> OFF 또는 0 Start Operating
ex1) keyl on
Operation
ex2) keyl off
Allow Panel
KEYLock[:STATe]?
Command that checks current power supply's front panel key & encorder switch adoptance.
Return value " 0 "
Allow panel condition
" 1 "
Stopped panel condition
ex) keyl?
return value "1"
LASTPower[:STATe] {SAFety|REMain|UPDate}
Command that restore previous output & panel condition.
> SAFety Operate *RST Command.
> REMain Maintain latest update.
> UPDate Restore setting just before OFF.
Note
Update command also restore OVP, OCP, LOCK, Output range, Sound function.
LASTPower[:STATe]?
Command that checks last power condition in current power supply.
Return value "SAF"
Safety Condition.
"REM"
Remain Condition.
"UPD"
U d t C diti
"UPD"
Update Condition.
ex) last?
return value "1"
*IDN?
Command that checks power supply's option.
Return value "ODA Technologies,OPC-3010,1.0-1.0-1.0"
1 Com
p
an
y
name
1
Company
name
2 Name of the model
3 Product detail version info.
1 System controller Version
2 Front panel Version
3 SCPI protocol Version
ex) *idn?
return value "ODA Technologies,OPC-3010,1.0-1.0-1.0"
SAV {1|2|3|4|5|6|7|8|10}
*SAV
{
1
|
2
|
3
|
4
|
5
|
6
|
7
|
8
|
10
}
Command that saves
p
ower su
pp
ly's voltage, current, OVP, OCP level in non-vertile memory in
1 ~ 10 memories.
> 1 ~ 10 Memory saving range
ex) *sav 2
save in to slot 2
- 73 -