EasyManua.ls Logo

OHAUS AP110 - RS232 Command Table

OHAUS AP110
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
hhmmsshhmmss
hhmmsshhmmss
hhmmss
TIME TIME
TIME TIME
TIME Set current time to "hh: mm: ss", hh is between 0 - 23.
TIMETIME
TIMETIME
TIME Print current time.
mmddyy DATEmmddyy DATE
mmddyy DATEmmddyy DATE
mmddyy DATE Set current date 'mmddyy'.
DATEDATE
DATEDATE
DATE Print current date.
%%
%%
% Print current % ref.
##
##
# Print current ref., any function.
ACAC
ACAC
AC Abort calibration /Test.
CC
CC
C Start an auto calibration.
EE
EE
E Go to initial state of current function.
FF
FF
F Print current function.
MM
MM
M Same as mode button.
xDxD
xDxD
xD Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON).
IDID
IDID
ID Print current ID string.
xxxxIDxxxxID
xxxxIDxxxxID
xxxxID Enter ID string (xxxx limit to 8 characters).
RS232 COMMAND TABLERS232 COMMAND TABLE
RS232 COMMAND TABLERS232 COMMAND TABLE
RS232 COMMAND TABLE
Field: Mode Stab CR LF
Length: 5 1 1 1
blank if stable
“ ? ” if unstable
CommandCommand
CommandCommand
Command
CharacterCharacter
CharacterCharacter
Character
DescriptionDescription
DescriptionDescription
Description
? ?
? ?
?
Print current unit.
0 = minimum level
1=
2 = maximum level
xIxI
xIxI
xI Set averaging level to "x", where x = 0, 1,
2
GRAMS, DWT, CARAT, OZ AV, OZ T, GRAIN,
TAEL 1, TAEL 2, TAEL 3, MOMME,CUSTOM UNIT
xMxM
xMxM
xM Places balance in mode "x",
where x = 1 to 11 (see table).
If unit is not already enabled,
command will be ignored.
1 = grams
2 = pennyweight
3 = carats
4 = avoidupois ounces
5 = troy ounces
6 = grains
7 = taels
8 = mommes
11 = custom unit
PP
PP
P Print display data
When “numeric only” display
data is selected for output in
the RS232 menu, the Mode
field is not output.
Field: Weight Mode Stab CR LF
Length: 10 1 5 1 1 1
Same as ?
command
Displayed weight sent right justified
w/lead zero blanking.
Nine characters include:
decimal point (1)
weight (7 max))
polarity (1): blank if positive
“ - ” if negative
INSTALLATION

Table of Contents