COMPUTER
INTERFACE
CONTROL
MODE
Computer
Communication
with
Ohmeda
Biox
3700
Oximeter
CONTROL
MODE
EXAMPLE
1118-301
In
order
to
change
the
Pulse
Volume,
press
in
succession ESCAPE CHI ENTER
The
H
is
the
capital
letter
which
represents
the
Pulse
Volume
and
the
1
(one)
is
the
parameter
to
which
you
want
to set the
Pulse
Volume.
If
the
information
has
been
entered
correctly,
the
Pulse
Volume
should
change.
If
the
information
has
been
entered
incorrectly,
a
WHAT?
appears
on
the
computer
screen
and
a
beep
is
heard.
If
the
Pulse
Volume
has
not
gone
to
one:
*
press
the
CAPS
LOCK
key
* press ESCAPE CHI ENTER
Entering
and
Enabling
The
following
information
is
important
to
know
when
using
the
Control
Mode:
Parameters
To
set
an
alarm
limit
or
pulse
volume
to
off,
input
0
(zero)
as
the
parameter.
If
no
parameter
needs
to
be
input,
press
ENTER
after
the capital letter entered.
The % and BPM do NOT need to be entered with the
parameter
for
Sa02
and
Pulse
Rate
limits.
The
Sa02
alarm
limits
change
in
steps
of
one.
The
Pulse
Rate
alarm
limits
change
in
steps
of
five.
The
message
WHAT?
appears
on
the
computer
screen
along
with an audible beep
if:
*
the
Oximeter
does
not
recognize
a
letter
or
parameter
field
* or a
parameter
has been
omitted.
WHAT?
05/09/86
82b