EasyManua.ls Logo

Oldham iTrans 2

Oldham iTrans 2
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
82
3
Unsigned-8
RTC Month Channel 1
4
Unsigned-8
RTC Year Channel 1
5-8
Unsigned-32
Total Operation Time (In Minutes) Channel 1
9
Unsigned-8
RTC Minute Channel 2
10
Unsigned-8
RTC Hour Channel 2
11
Unsigned-8
RTC Day Channel 2
12
Unsigned-8
RTC Month Channel 2
13
Unsigned-8
RTC Year Channel 2
14-17
Unsigned-32
Total Operation Time (In Minutes) Channel 2
Command 131 Read User Configuration Response Length: 36 Bytes
Byte Number
Parsing
Parameter
0-3
Float IEEE 754
Low Alarm Threshold Channel 1
4-7
Float IEEE 754
High Alarm Threshold Channel 1
8-11
Float IEEE 754
Analog Output Range Channel 1
12-15
Float IEEE 754
Cal Gas Value Channel 1
16-17
Unsigned-16
Calibration Interval In Days Channel 1
18-21
Float IEEE 754
Low Alarm Threshold Channel 2
22-25
Float IEEE 754
High Alarm Threshold Channel 2
26-29
Float IEEE 754
Analog Output Range Channel 2
30-33
Float IEEE 754
Cal Gas Value Channel 2
34-35
Unsigned-16
Calibration Interval In Days Channel 2
Command 132 Read Live Channels Information Response Length: 26 Bytes
Byte Number
Parsing
Parameter
0-3
Float IEEE 754
User Peak Channel 1
4-5
Unsigned-16
Previous OR Channel 1
6
Unsigned-8
Last Alarm Day Channel 1
7
Unsigned-8
Last Alarm Month Channel 1
8
Unsigned-8
Last Alarm Year Channel 1
9-10
Unsigned-16
Max Temperature Channel 1
11-12
Unsigned-16
Min Temperature Channel 1

Table of Contents