EasyManua.ls Logo

Oldham MX 32 - Page 43

Oldham MX 32
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Menus
and sensitivity or Esc to exit the procedure.
2. The detector is calibrated.
5. Sensor exchange
This function reboots the parameters (rate of wear, calibration date, internal
parameters corresponding to the 4-20mA range, etc.) from the selected
detector(s) following or in view of a change of cell.
Detector Selection
1. Select the detector(s) to be rebooted with the help of the Previous
detector, and Next detector keys and press Select.
Initialization of the Detector
1. Press Escape to reset the selected cells
2. Proceed with the cell replacement and then calibrate locally the
corresponding detectors
3. Proceed with the calibration from the MX 32 via the menus1 Sel
detectors”, “2 recording”, "End recording” and “4 validation in order to
save data calibration (wear rate, date of calibration, response time, etc.).

Table of Contents

Other manuals for Oldham MX 32

Related product manuals