9 Cleaning and maintenance
Chapter 9 │Cleaning and maintenance
Cleaning
Do not use alcohol- or ammonia-based liquids to clean the controller. If
necessary, clean the exterior of the enclosure with a damp cloth.
Fuse replacement
Fuse replacement should only be performed by qualified
personnel and power must be first switched off.
Fuses shall comply with IEC 60127 standard (time-delay fuse, low breaking
capacity, 250Vac). Please see Chapter 8.
Replacement of the lithium battery
Lithium battery replacement should only be performed by qualified personnel
and with an identical battery (see Chapter 8 for the list of spare parts). The
controller must be first switched off. Power on the controller once the battery
has been replaced.
Oldham does not allow any other repairs than those listed here
above.
Burn hazard. As the temperature inside the controller can reach
70°C (158 °F), it should be allowed to cool after aperture.