CompleteTeardown,Cleaning,andReassemblyoftheOlympusBH‐2SlidingFocusBlock
Revision2 Page18of23
ReinstalltheArmontotheBase
Placethe arm flat on its back onthe worksurfacewith
thebottomfacingyou(seeFigure55).
Figure55–Placethearmflatontheworksurface
Holdthebaseofthestandinpositionsuchthatthefour
holesinthebasealignwiththefourtappedholesinthe
bottomsurfaceofthearmcasting(seeFigure56).
Figure56–Alignholesinthebasewithholesinthearm
Use a 5mm Allen wrench or driver to loosely reinstall
four M6x25 hex, socket‐head cap screws to attach the
armtothebaseofthestand(seeFigure57).
Figure57–Attachthebaseontothearmofthestand
Check the alignment of the arm (relative to the base)
and adjust the positioning as necessary for an
acceptable cosmetic appearance at the interface.
Tightenthe fourM6x25screwstosecurethearmtothe
base(seeFigure58).
Figure58–Securethearmontothebaseofthestand
ReinstalltheElectricalBase
Carefullylower theassembled stand overthe electrical
base, maintaining proper alignment relative to the
electrical base (to prevent damage to the optical
components in thebase)until the standseats ontothe
base.If the electrical base contains a light‐preset
control/switch (see Figure 5), be sure to reengage
this
into the notch in the side of the base
3
before seating
the stand onto the electrical base.While holding the
electricalbase inplace so thatitdoes notshift, lay the
standonto itsback orside onthe worksurface. Usea
4mm (or 5/32”) Allen wrench ordriver toreinstall four
M5x12 hex, socket‐hea d cap
screws and associated
washers to secure the electrical base to the stand (see
Figure 59).Return the assembled stand to its normal
uprightposition.
Figure59–Reinstallscrewstosecuretheelectricalbase
3
Orreconnectthetwoelectricalconnectorstotheprintedcircuitboard.