Flam",a',le gases
CD
The CLV-U20 light source should be used only
in
a medical facility
under
the
supervision
of
a
trained physician.
@ The light source is designed (floating)
to
function
integrally
with
insulated patient
attachement
(endoscopes)
to
prevent operator/patient shock. Endoscopes and cameras
to
be used
with
it
should
not
be grounded
in
order
to
prevent
shocks
which
may
be caused
by
leakage
current
from other electrical apparatus applied
to
the
patient
The endoscopist should wear rubber
gloves as an added precaution.
® The housing of the light source must
be
grounded securely and effectively. Do not defect
the
line cord ground connection.
@)
The light source/endoscope combination should never be applied
to
the
heart directly.
@ Never install and operate the light source where there
is
the risk
of
flammable gases.
@ Keep liquids away from
all
electrical equipment
to
prevent operator shock and instrument
damage. Do
not
use
the
light source if spilled fluids have entered
the
unit.
~
~\,
1\
"
,
Powe'
PIU:~~
Hospital
grade
receptacle
(wall
mains
outlet)
Be
sure
to
connect
the
power
plug
to
a
hospital
grade
receptacle
(a
wall
mains
outlet).
l_l_-3_S_a--'-rfi_
e
_ty_P--'-_re_c_a_u_tio_n_s
-----)::===========
Spillable
liquids