Using Software Options
Part # 910-250C
109
Figure 59 Intermediate Echo Setup
Once setup is complete, the operator inspects unknown welds. When a ring-down pattern
appears from a weld, the Spotweld Assistant displays a weld rating in the upper right
corner of the live A-Scan that is continuously updated with the readings on the screen.
When weld inspections are stored to the datalogger, this rating is also saved as a file tag,
and can be either accepted or rejected by the trained operator as a correct weld rating. This
dynamic rating is a recommendation only and should not substitute for the judgment of a
trained professional inspector.
9.8.3 Option Activation
Prior to activating the EPOCH 4PLUS Spotweld Assistant option, the operator should
properly calibrate the instrument for the transducer and material being inspected. The
Spotweld Assistant option is located in the OPTION Menu, under Spotweld. To activate
the Spotweld Assistant, follow the steps below:
1. Highlight Spotweld from the OPTION Menu and press [ENTER].
2. Select ASSISTANT from the Spotweld Menu and press [ENTER].
3. Use the slewing keys to select ON from the list and press [ENTER] to turn the
function on.
4. To complete activation, press [OPTION] to exit the menu and begin using the
Spotweld Assistant.