EPOCH 4PLUS
4.4 Using the Pulser Key
The following pulser receiver and filter parameters can be adjusted on the EPOCH 4PLUS
keypad:
Waveform Rectification
Pulser Energy
•Pulser Damping
Pulser Test Mode
Filter Selection
Pulser Type (square wave or spike excitation)
Pulser Frequency (tuned for square wave pulser only)
To access the available parameters, follow these steps:
1. Press [PULSER] multiple times to toggle through the several features.
2. Press [2nd F], [PULSER] (FILTER) to go directly to the filter parameter.
Each pulser-receiver and filter characteristic adjustments can be done in either full A-
Scan or Split Screen.
In the full A-Scan, the current parameter being adjusted appears at the top left of
the screen and directly above the waveform.
In the Split Screen mode, the parameter is highlighted on the lower half of the
EPOCH 4PLUS screen.
Note: For reviewing the adjustments in this manual, the Split Screen mode was
used.
To view how the selections of the pulser receiver affect a waveform, couple a transducer
to a sample or test block, which produces an echo on the top half of the screen.
4.4.1 Waveform Rectification
The EPOCH 4PLUS can operate in one of four different rectification modes:
Full Wave
Half Wave Positive
Half Wave Negative
RF (unrectified)
The RF mode is not active while operating in DAC or TVG mode, or Peak Memory.
To select a waveform rectification mode, follow these steps:
1. Press [PULSER] once to display the current screen rectification. This is highlighted
and displayed at the bottom of the screen.