Olympus Panametrics Epoch 4 Plus - 9.10.3 Operating in -12 db Mode

Olympus Panametrics Epoch 4 Plus
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using Software Options
Part # 910-250C
127
Figure 78 Example of -6 dB Mode
9.10.3 Operating in -12 db Mode
In –12 dB mode, the desired gate floats at 12 dB below the peak echo amplitude in the gate.
This corresponds to 25% of the echo’s maximum height. The screen below depicts Gate 1
with Floating Gate active in the –12 dB mode. Note that the gate is exactly at 25% of the
echo’s peak amplitude.
Note: The Floating Gate option is not valid in RF mode. Also, Floating Gate
cannot be turned on or off while the screen is frozen.
Figure 79 Example of -12 dB Mode