Using Software Options
Part # 910-250C
145
Figure 97 Snowflake Icon
Figure 98 Standard FREEZE “F” Icon
When the EPOCH 4PLUS Auto-Freeze is triggered in HOLD MODE, the standard
FREEZE “F” icon appears. If the EPOCH 4PLUS Auto-Freeze is triggered in TIME
MODE, the “F” icon will flash for the duration of the freeze.
Note: This option does not affect the EPOCH 4PLUS’s datalogger or the
GageView PRO Interface Program. The Auto-Freeze option is not saved
with an instrument calibration file. When a calibration is recalled, the Auto-
Freeze option must be activated separately.
When a gate alarm is triggered in Auto-Freeze mode, the EPOCH 4PLUS
will beep for one second.
The FREEZE key is active in Auto-Freeze TIME MODE. If the user presses
the FREEZE key during a timed freeze, the gage immediately unfreezes.
This screen shows that Auto-Freeze has
been activated but that it has not been
triggered. The Snowflake Icon has
appeared after pressing 2
nd
F FREEZE.
In order to trigger Auto-Freeze, a Gate
Alarm must be triggered.
This screen shows the EPOCH 4PLUS
when Auto-Freeze has been triggered.
Notice that the standard FREEZE “F” icon
has replaced the Snowflake Icon. The
Auto-Freeze was triggered due to the
active positive logic alarm on Gate 1.