3.3 Connecting a CO2 gas cylinder
27
UCR INSTRUCTION MANUAL
Ch.3
• If a significant gas leak is noticed from within the endoscopic CO
2
regulation unit,
stop using the endoscopic CO
2
regulation unit immediately and contact Olympus.
Attach the CO
2
cylinder holder (MAJ-188, MAJ-1614) to the Olympus mobile
workstation (WM-NP1 or WM-WP1), and attach the CO
2
cylinder to the holder.
The cylinder hose must be attached to the endoscopic CO
2
regulation unit by Olympus personnel
during installation. The enclosed nut cover will be fixed during the installation when the cylinder hose
and the endoscopic CO
2
regulation unit are connected by Olympus personnel.
1 Inspect the cylinder hose for the endoscopic CO
2
regulation unit for damage, cracks,
and other irregularities.
2 Set the gas source selection switch on the rear panel to the “Cylinder” mode.
3 Use the supplied wrench to attach the cylinder
hose to the CO
2
gas inlet on the rear panel of
the endoscopic CO
2
regulation unit and
securely tighten it with a force of about
24.5 N·m (2.5 kgf/m) clockwise.
Figure 3.1