28
3.3 Connecting a CO2 gas cylinder
UCR INSTRUCTION MANUAL
Ch.3
4 When using the cylinder hose (PIN, MAJ-1080), attach the clamp to a gas cylinder
filled with CO
2
gas. Attach the clamp to the adapter by placing the pin of the clamp
into the guide hole of the cylinder, and tighten the handle with a force of about
17.2 N·m (1.8 kgf/m).
Figure 3.2
5 When using the cylinder hose (MAJ-1081, MAJ-1082), use the supplied wrench to
attach the cylinder hose's adapter to a gas cylinder filled with CO
2
gas.
Figure 3.3
6 Confirm that the endoscopic CO
2
regulation unit and the CO
2
gas cylinder are
correctly connected and open the gas cylinder valve by turning it counterclockwise.
Figure 3.4
CO
2
gas
cylinder
Clamp
Cylinder hose
Pin
Handle
Adapter
MAJ-1081 (DIN), MAJ-1082 (ISO)
CO
2
gas
cylinder
Handle
Gas cylinder
valve