6.5 Reprocessing the endoscopic CO2 regulation unit and accessories
UCR INSTRUCTION MANUAL
Ch.6
Disinfection
Follow the usage (temperature, contact time, use life) and amount (concentration)
provided by surface disinfectant cleaner manufacturer. Failure to follow
manufacturer's instructions may result in insufficient cleaning and disinfection
effect.
• Wipe off remaining surface disinfectant cleaner solution according to instructions of
surface disinfectant cleaner manufacturer. Failure to do so may adversely affect the
human body or this instrument.
• After wiping with a moistened lint-free cloth, dry the endoscopic CO
2
regulation unit
and accessories thoroughly before using it again. If it is used while still wet, there is
the risk of an electric shock.
1 Turn OFF the endoscopic CO
2
regulation unit and disconnect the power cord from the
hospital grade wall mains outlet.
2 Prepare a lint-free cloth moistened with the
surface disinfectant cleaner.
Figure 6.1
3 Wipe off all patient debris from the endoscopic CO
2
regulation unit and accessories
this instrument using the moistened lint-free cloth.
Lint-free cloth moistened with
the surface disinfectant cleaner