6.5 Reprocessing the endoscopic CO2 regulation unit and accessories
61
UCR INSTRUCTION MANUAL
Ch.6
Cleaning
After wiping with a moistened lint-free cloth, dry the endoscopic CO
2
regulation unit
thoroughly before using it again. If it is used while still wet, there is the risk of an
electric shock.
4 Prepare another lint-free cloth moistened with
the surface disinfectant cleaner.
Figure 6.2
5 Disinfect the surfaces that previously contained patient debris (see Step 3) by wiping
with the moistened lint-free cloth.
6 Ensure that the surfaces are completely wet for the contact time instructed by the
surface disinfectant cleaner manufacturer.
7 If any surface of the endoscopic CO
2
regulation unit and accessories remains wet,
wipe it with a dry lint-free cloth and let it dry thoroughly.
1 Turn the endoscopic CO
2
regulation unit OFF and disconnect the power cord from the
hospital grade wall mains outlet.
2 Wipe with a dry lint-free cloth or a moistened lint-free cloth with water until dust and
dirt are removed.
3 If the surface of the instruments is wet, wipe it with a dry lint-free cloth and let it dry
thoroughly.
Lint-free cloth moistened with
the surface disinfectant cleane
r