DP20 Economical Multisignal Meter Instruction Manual
1.29 Conguraon menu (cont.)
Password
Version
Minimum
Brightness
Maximum
Standard
‘Eco’ mode
Degrees
Degrees Celsius
Degrees Fahrenheit
Temperature
resoluon
1º resoluon
0.1º resoluon
Alpha
Alpha 0.0385
Alpha 0.0390
the second scaling (ScL.2’) funcon allows to congure the
reading associated to the ‘second scaling, which can be acvated
by controlling a contact at the external control’. For more
informaon on the ‘second scaling’ funcon see secon 1.24.
the Eco mode(‘Eco’) is a funcon to reduce consumpon of
the instrument, by powering o the display when display is not
needed. For consumpon data in ‘Eco’ mode see secon 1.8. For
more informaon in ‘Eco’ mode see secon 1.22. The operaon
of the ‘Eco’ mode is explained below :
- the waing me unl display is powered o display is
congurable from 5 to 255 seconds.
- aer the congured me is over, without interacon from
the operator, the display powers o the reading. It will not
power o if the instrument is into conguraon menu’ or in
‘fast access’ menu or there is an acve alarm.
- the instrument will power on the display when an alarm
acvates (either alarm 1 or 2) and will remain on if alarm
remains acve.
- the instrument will power on the display if the operator
press any front key.
at the Temperature resoluon (‘t.rES’) menu select the
resoluon to or 0.1º. This resoluon applies to temperature
probes Pt100, Pt500, Pt100, Ni100, Ni200, Ni1000, PTC and NTC.
For thermocouple probes resoluon is xed to 1º.
at the ‘Degrees’ (‘dEG’) menu select the temperature values to
be read in ºC (degrees Celsius) or in ºF (degrees Fahrenheit).
at the Alpha (‘ALPh’) menu select the value for the alpha
parameter to ‘0.0385’ o ‘0.0390’. This parameter aects to Pt
probes (Pt100, Pt500, Pt1000) and it depends on the probe.
at the Cold juncon (‘cJc’) menu select on’ to acvate the
thermocouple cold juncon compensaon. Select o to
deacvate.
the AC deadband (‘d.bnd’) parameter applies only to
AC measuring ranges. The True RMS measures deal with a
considerable noise when signal is near to 0. This parameter allows
to set a value between 0 and 100 (by default is 20) to empirically
set the signal level at which reading will be forced to 0, when
signal is below that level.
at the Brightness (‘LIGh’) menu select between 5 levels of
brightness intensity for the display. It adapts the instrument to
environments with higher or lower brightness is needed or to
adapt the intensity to other meters in the area.
the Version(‘VEr’) menu informs about the rmware version
installed on the instrument.
at the Password(‘PASS’) menu select a 4 digit code to block
access to the conguraon menu. Use the Passwordfuncon
to prevent modicaons on the instrument conguraon. To
acvate the Passwordselect ‘onand enter the numerical code.
The code will be requested when trying to access the conguraon
menu’. The ‘fast access’ menu is not blocked by the ‘Password’.
At the Factory reset (‘FAct’) menu, select yes to load the
Second scaling
Decimal point
Display Low
Display High
5 to 255 seconds
AC deadband’
Deadband for AC signals
Cold juncon
Thermocouple cold
juncon compensaon
‘on’ or ‘o