• the ‘second scaling’ (ScL.2’) funcon allows to congure the
reading associated to the ‘second scaling’, which can be acvated
by controlling a contact at the ‘external control’. For more
informaon on the ‘second scaling’ funcon see secon 1.24.
• the ‘Eco mode’ (‘Eco’) is a funcon to reduce consumpon of
the instrument, by powering o the display when display is not
needed. For consumpon data in ‘Eco’ mode see secon 1.8. For
more informaon in ‘Eco’ mode see secon 1.22. The operaon
of the ‘Eco’ mode is explained below :
- the waing me unl display is powered o display is
congurable from 5 to 255 seconds.
- aer the congured me is over, without interacon from
the operator, the display powers o the reading. It will not
power o if the instrument is into ‘conguraon menu’ or in
‘fast access’ menu or there is an acve alarm.
- the instrument will power on the display when an alarm
acvates (either alarm 1 or 2) and will remain on if alarm
remains acve.
- the instrument will power on the display if the operator
press any front key.
• at the ‘Temperature resoluon’ (‘t.rES’) menu select the
resoluon to 1º or 0.1º. This resoluon applies to temperature
probes Pt100, Pt500, Pt100, Ni100, Ni200, Ni1000, PTC and NTC.
For thermocouple probes resoluon is xed to 1º.
• at the ‘Degrees’ (‘dEG’) menu select the temperature values to
be read in ºC (degrees Celsius) or in ºF (degrees Fahrenheit).
• at the ‘Alpha’ (‘ALPh’) menu select the value for the alpha
parameter to ‘0.0385’ o ‘0.0390’. This parameter aects to Pt
probes (Pt100, Pt500, Pt1000) and it depends on the probe.
• at the ‘Cold juncon’ (‘cJc’) menu select ‘on’ to acvate the
thermocouple cold juncon compensaon. Select ‘o’ to
deacvate.
• the ‘AC deadband’ (‘d.bnd’) parameter applies only to
AC measuring ranges. The True RMS measures deal with a
considerable noise when signal is near to 0. This parameter allows
to set a value between 0 and 100 (by default is 20) to empirically
set the signal level at which reading will be forced to 0, when
signal is below that level.
• at the ‘Brightness’ (‘LIGh’) menu select between 5 levels of
brightness intensity for the display. It adapts the instrument to
environments with higher or lower brightness is needed or to
adapt the intensity to other meters in the area.
• the ‘Version’ (‘VEr’) menu informs about the rmware version
installed on the instrument.
• at the ‘Password’ (‘PASS’) menu select a 4 digit code to block
access to the ‘conguraon menu’. Use the ‘Password’ funcon
to prevent modicaons on the instrument conguraon. To
acvate the ‘Password’ select ‘on’ and enter the numerical code.