Test Set CMC 156
44
5.5.2 Shock and Vibration
Dynamics
Vibration Tested according to IEC68-2-6 (operating mode)
frequency range 10 to 150 Hz; acceleration 2 g
continuous (20 m/s
2
); 10 sweeps, each axis.
Shock Tested according to IEC68-2-27 (operating mode)
15 g/11 ms, half-sine, each axis.
Table 25 Shock and Vibration
5.5.3 Electromagnetic Compatibility (EMC)
EMC
CE conformity The product complies to the normative document
about electromagnetic compatibility for
standardization of the laws of the member states of
the council of the European Union (EMC standard
89/336/EEC).
Emission
International
Europe
EN 50081-2
FCC Subpart B of Part 15 Class A
Susceptibility
International
Europe
EN 50082-2:1992
IEC 1000-4-2/3/4/6
Table 26 Electromagnetic Compatibility