-82 -
Autonomous operation also works via the industrial/stackable PIF of the Xi 80. The device is
powered by the power supply of the PIF.
Up to 9 measure areas (for Xi 80 with firmware 3013 or higher) and up to 3 measure areas
(Xi 410) can be output autonomously. The use of three stackable PIFs is required. Three
analog outputs are possible per stackable PIF. The response time is 20 ms for Xi 80 and
640 ms for Xi 410.
Note: Several measuring areas can be combined into one measuring area via a so-called
super area.
Autonomous mode cannot be set using the Main measure area function (configuration menu-
Device (PIF)). To output a measure area autonomously, the function Measure area must be
used.