SAMPLE ANALYSIS - CONCENTRATES
3. Gather at least 100mg of sample. Most samples do not need any specific
preparation; however, this may vary depending on the sample type.
4. Place scale on level surface and turn on.
5. Check calibration of scale by placing 10g calibration weight on scale. Results
should be between 0.997 and 10.003. If not proceed to Section 4 to recalibrate
scale.