SAMPLE ANALYSIS - CONCENTRATES
6. Place a vial cap on the scale and press “tare”. Ensure scale now reads zero with
cap resting on it. Press forward arrow on the screen.
7. Add 100mg (+/- 10mg) to the vial cap. Make sure all sample is within inner
circle of cap. Press forward arrow on the screen.
8. Type in the exact sample reading on the scale. 90-110mg is an acceptable range.
Press forward arrow on the screen. Note 100mg is 0.100g.