SAMPLE ANALYSIS - CONCENTRATES
17. Turn on shaker and press Start on screen.
18. Zero process will begin, which takes two minutes to complete.
19. After completion of the zero, inspect the vial. If the sample has not fully
dissolved or if there is a significant amount of residue on the vial from the
sample repeat the heating and extraction process again. Do not continue until all
the sample is dissolved or the results may not be accurate.
20. Once Zero is complete, add syringe filter onto sample port. Press forward arrow
on screen.