SAMPLE ANALYSIS - CONCENTRATES
14. Place the extraction vial into the sample heater, then fill the included measuring
cup with water and pour contents into the heater. Most samples will be heated
sufficiently by filling measuring cup to the “Eline. Press firmly down on the
button near the bottom of the heater to start heating.
15. The heater will boil water and make steam which causes the contents of the
sample vial to warm up. Heating takes approximately 2 minutes.
16. Once the orange light on the heater turns off and all water in the heater has
boiled away, remove vial carefully (it may be hot!) and place vial into shaker and
strap in place. Press forward arrow on screen. NOTE: For best results place
sample into shaker immediately after heating is complete. If the sample cools off
the extraction may be poor.