SAMPLE ANALYISIS- WET FLOWER
22. Place syringe on syringe filter at sample port by gently twisting it on about a
quarter turn. Press forward arrow on screen.
23. Slowly inject at least 1ml of sample into sample port. Exact amount isn’t
important. There will be some resistance when injecting a sample- this is
normal. Press forward arrow on screen.
24. Set valve to Run. Press forward arrow on screen.