SAMPLE ANALYISIS- WET FLOWER
25. Press Start. The LightLab will begin to analyze the sample. Results will be ready
in 8 minutes.
26. Once sampling is complete, results are displayed on the screen. A note can be
added to the sample if desired. If SD card is installed, results will be added to a
comma separated value (Excel compatible) file. Do not reuse syringe, filter or
vial for future tests unless re-running the same sample.
Note: Detection limits will vary depending on the options selected using the
following equation:
Detection Limit (%)= (100/(Weight Entered))*(ml Solvent Added)/10)*(1-
100/(Moisture Content Entered))
In general terms:
- An increase in sample weight will provide a lower detection limit
- An increase in solvent volume will provide a higher detection limit
- An increase in moisture content will provide a lower detection limit