SAMPLE ANALYISIS- RAFFINATE
6. Place scale on level surface and turn on.
7. Check calibration of scale by placing 10g calibration weight on scale. Results
should be between 0.997 and 10.003. If not proceed to Section 4 to recalibrate
scale.
8. Place a vial cap on the scale and press “tare”. Ensure scale now reads zero with
cap resting on it. Press forward arrow on the LightLab screen.