SAMPLE ANALYISIS- RAFFINATE
9. Add 200mg (+/- 20mg) to the scale. Make sure all sample is within inner circle
of cap. Press forward arrow on the screen. NOTE: if moisture content was set to
a value greater than 25%, add 500mg (+/- 50mg) to the scale.
The sample may alternatively be weighed in a small weigh boat or the sample
vial. If a weigh boat is used, we recommend placing the whole weigh boat into
the sample vial so that no trichomes are lost during the transfer to the vial.
10. Type in the exact sample reading on the scale. 180-220mg is an acceptable range.
Press forward arrow on the screen. Note 100mg is 0.100g.
11. Pour sample into the vial. Ensure all the sample is inside the vial and not
trapped in the vial cap threads. Press forward arrow on the LightLab screen.