9.6 Collecting Head Impulse data
For information about setting various Head Impulse options, refer to the Reference Manual.
Note The temporary patient is available to allow testing to start without entering any patient data. Refer to Patient
data entry/lists 7 to create a new patient or open an existing patient.
1. The Impulses settings (A) define how many valid left and right impulses are required before the test stops auto-
matically.
To change the number of suggested minimum head impulses for the current testing session, click the up or down
arrows or type in the number directly. Twenty (20) is the recommended minimum number of head impulses for both
leftward and rightward impulses.
Note Impulse settings can be any number up to 999. For information about setting the default numbers, refer to
the Reference Manual.
54 Otometrics - ICS Impulse USB
9 Head Impulse