EasyManua.ls Logo

Owon SDS7202 - Alternate Trigger

Owon SDS7202
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.Advanced User Guidebook
40
Mode
Holdoff
Auto
Normal
Single
Holdoff
Reset
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust M knob to set time interval
before another trigger occur.
Set Holdoff time as 100 ns
Alternate trigger
Trigger signal comes from two vertical channels when alternate trigger is on. This
mode is used to observe two unrelated signals. You can choose different trigger modes
for different channels. The options are as follows: edge, video, pulse or slope.
1. Alternate trigger (Trigger mode: Edge)
Alternate trigger (Trigger Type: Edge) Menu is shown as Figure 5-23.
Figure 5-23 Alternate trigger (Trigger Type: Edge) Menu
Alternate trigger (Trigger Type: Edge) Menu list:
MENU
SETTING
INSTRUCTION
Alternate Mode
Edge
Set vertical channel trigger type as edge trigger.
Source
CH1
CH2
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Coupling
AC
DC
HF
LF
Not allow DC portion to pass.
Allow all portion pass.
Not allow high frequency of signal pass and only
low frequency portion pass.
Not allow low frequency of signal pass and only
high frequency portion pass.
Slope
Trigger in signal rising edge
Trigger in signal falling edge
Mode
Holdoff
Auto
Holdoff
Reset
Acquire waveform even no trigger occurred
100 ns - 10 s, adjust M knob to set time interval
before another trigger occur.
Set Holdoff time as 100 ns

Table of Contents

Related product manuals