Oxitone 1000M
User Manual Rev. 1.
Page 35 of 39
Table 1: Electromagnetic Emission
Emission Test Compliance Electromagnetic Environment - Guidance
This device is intended for use in the electromagnetic environment specified below:
The customer and / or user of this should ensure that it is used in such an environment
RF Emissions
CISPR11
Group 1 This device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF Emissions
CISPR11
Group B This device is suitable for use in all
establishments, including domestic and
those directly connected to the public low-
voltage power supply network that supplies
building s used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
N / A
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
N / A