Table 2: Electromagnetic Immunity
Immunity Test IEC Test Level
60601-1-2
Compliance
Level
Electromagnetic
Environment Guidance
This device is intended for use in the electromagnetic environment specified below:The
customer and / or user of this should ensure that it is used in such an environment
Electrostatic
Discharge (ESD)
IEC 61000-4-2
8 kV contact
± 15 kV air discharge
± 8 kV contact
± 15 kV air
discharge
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material,
relative humidity should
be at least 30%.
Electrical Fast
Transient / Burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for power supply
lines
N / A N / A
Surge
IEC 61000-4-5
± 1 kV different mode
± 2 kV common mode
N / A N / A
Voltage dips,
short
interruption, and
voltage
variations on
power supply
input lines
IEC 61000-4-11
± 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
± 40% U
T
(60% dip in U
T
)
for 5 cycles
± 70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 sec.
N / A N / A
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial,
hospital or home
environment.