Oxitone 1000M
User Manual Rev. 1.
Page 37 of 39
NOTE: U
T
is the AC mains voltage before application of the test level.
Table 3: Guidance and Manufacturer's Declaration-Electromagnetic Immunity
Immunity Test IEC 60601-1-2 Test Level
Compliance
Level
Electromagnetic
Environment Guidance
This device is intended for use in the electromagnetic environment specified below:
The customer and / or user of this should ensure that it is used in such an environment
Portable RF communications equipment should be used no closer to any part of the device,
including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3V
rms
150 kHz to 80 MHz
10V/m
80 MHz to 2.5 GHz
N / A
10V/m
Recommended Separation Distance
N / A
80 MHz to 800 MHz 𝑑 = 1.17√𝑃
800MHz to 2.5 GHz d = 2.33 √𝑃
Where P is the maximum output power
rating of the transmitter in wats (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol: