5.1 Setup: Scale
Scale Manuf: Select the appropriate value for the scale in use. Options: Ohaus, Ohaus2,
Ohaus3, Mettler, Metler2, Sartor, Sartor2. Default = Ohaus3. Proper configuration of the
scales parameters is required, as well as correct interface cable.
Ohaus3: Adventurer Pro.
Ohaus2: Adventurer, Explorer, Explorer Pro.
Mettler: Viper, Series 4, IND560.
Metler2: Speedweigh, Panther.
Sartor2: Current default Sartorius setting, all series.
Alarm: Triggered if communication with the scale is lost. Options: Pump Stop, Alarm Only
or Off. Default = Pump Stop.
Tare Weight: Determines if the system tares the scale upon pressing Execute and Run in
the main operational modes. This is counterintuitive. Options: On, Off. Default = OFF, which
causes the system to tare the scale. ON will cause the tare to not occur.
Units: Select from: .001 gm, .01 gm, 0.1 gm, Kg, T, Lbs, OzT, Oz, C, Dwt. Default = .01
gm.
- SETUP - SCALE
UP DOWN SELECT
- SCALE- SCALE MANF.
UP DOWN SELECT
SELECT
SCALE MANF: OHAUS3
UP DOWN SELECT
- SCALE- ALARM
UP DOWN SELECT
- SCALE- TARE WEIGHT
UP DOWN SELECT
- SCALE- UNITS.
UP DOWN SELECT
ALARM: PUMP STOP
UP DOWN SELECT
TARE WEIGHT: OFF
UP DOWN SELECT
UNITS: .01 GM
UP DOWN SELECT
DOWN
DOWN
DOWN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Parker SciLog FilterTec Plus and is the answer not in the manual?

Parker SciLog FilterTec Plus Specifications

General IconGeneral
BrandParker
ModelSciLog FilterTec Plus
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals