1
SciLog
®
FilterTec Plus
®
Installation, Operation & Maintenance Instructions

Questions and Answers

  • M
    Michelle JonesAug 18, 2025
    Why is there no fluid flow from my Parker SciLog FilterTec Plus Industrial Equipment magnetic pump head, even though it's turning?
    • M
      merrittjenniferAug 18, 2025
      If your magnetic pump head runs but no fluid flows, ensure the pump stays wet, especially if it has run dry. Use a check valve as a dispensing tip to prevent fluid backflow. If you're using a high viscosity fluid, consider reducing the viscosity, applying head pressure, or opting for a different head style.
  • W
    wilsonelaineAug 20, 2025
    What to do if my Parker SciLog FilterTec Plus piston pump head turns, but no fluid flows?
    • S
      stephanie04Aug 20, 2025
      If your piston pump head runs but no fluid flows, it could be due to a broken piston or a loose coupler. Contact Parker for technical support or to arrange service and obtain an RGA#.
  • G
    George RodriguezAug 22, 2025
    How to fix a “Scale Error” on my Parker SciLog FilterTec Plus Industrial Equipment?
    • M
      Michael AllenAug 22, 2025
      To resolve a “Scale Error” message on your screen, press any key to clear the error. Then, tighten the cables and turn the scale on, and try again. If problems persist, verify the scale settings or contact Parker tech support for assistance, especially if you're not using the default scale.
  • T
    Terri GutierrezAug 26, 2025
    How to resolve “Check Pump Head” error on Parker SciLog FilterTec Plus Industrial Equipment peristaltic head?
    • T
      Timothy SextonAug 26, 2025
      If you encounter a “Check Pump Head” error with your peristaltic head, first ensure you are using the correct size tubing, as thick-walled tubing in a thin wall pump head can trigger this error. If the error occurs without tubing in the head, contact Parker, as you may need a new motor.
  • F
    frodriguezAug 27, 2025
    What to do if my Parker SciLog FilterTec Plus pump is acting weird?
    • M
      Madison BarberAug 28, 2025
      If the pump is behaving erratically with incorrect flows and pressures, or slowing down prematurely, this could be due to power spikes or brown outs. As a last resort, perform a Factory Reset by going to Setup; Pump, and then selecting Factory Reset to restore the unit to its default values.
  • K
    kkaiserAug 30, 2025
    Why is the pressure reading on my Parker SciLog FilterTec Plus always 0.0?
    • M
      matthewruizAug 30, 2025
      If the pressure reading is always 0.0, even when the sensor is disconnected, this usually happens when the sensor is Zeroed while not connected, or if Span is pressed instead of Zero. To fix this, you must perform a Factory Reset.
  • T
    Tracy MillerSep 1, 2025
    Why am I not getting data from the balance on my Parker SciLog FilterTec Plus Industrial Equipment?
    • K
      kristin17Sep 1, 2025
      If you are not getting data from the balance, make sure that the Filtrate Weight Alarm is enabled. Set this alarm to either Alarm Only, or Pump Stop.
  • K
    Katie BurtonSep 4, 2025
    How to fix “Device Error, Com Port Not Available” on Parker Industrial Equipment?
    • G
      George WrightSep 4, 2025
      If you receive a “Device Error, Com Port Not Available” error from your computer, it means the specified Com port is either in use or doesn't exist. Check Device Manager in your computer's properties, and expand the Ports (Com & LPT) section.
  • T
    Timothy JonesSep 7, 2025
    What causes “Check Pump Head” error on Parker SciLog FilterTec Plus piston or magnetic gear head?
    • P
      Pamela SpearsSep 7, 2025
      If you encounter a “Check Pump Head” error with your piston or magnetic gear head, you can try wetting the head by placing an appropriate solvent in the upper tubing overnight, if you believe it's stuck due to being dried out. If it still won’t turn, contact Parker to arrange an RGA to send your pump in for service, or purchasing a service kit if you have a magnetic gear head.
  • A
    ashley73Sep 9, 2025
    Why is my Parker SciLog FilterTec Plus peristaltic pump head turning, but no fluid flowing?
    • V
      Veronica MartinSep 9, 2025
      If the peristaltic pump head turns but no fluid flows, ensure you are using the correct size tubing. Tandem 1081 is for thin-walled tubing, and 1082 is for thick-walled tubing. Also, the pump head may be cracked from being forced closed with the tubing crosswise or the wrong size tubing; contact Parker for repair or replacement parts.