SciLog
®
FilterTec Plus
®
Installation, Operation & Maintenance Instructions

Questions and Answers

  • B
    Brian MooreSep 12, 2025
    What to do if Parker SciLog FilterTec Plus pump head turns without tubing, but not with tubing?
    • R
      rayjosephSep 12, 2025
      If the pump head turns ok with no tubing installed, won’t turn when you put tubing in, and you don’t get a “Check Pump Head” error, the coupler is likely loose or broken. Contact Parker for tech support, or to arrange for service and an RGA#.
  • U
    udavisSep 14, 2025
    Why is SciDoc not collecting data from FilterTec Plus on my Parker Industrial Equipment?
    • D
      Dr. Rebecca ChandlerSep 14, 2025
      If SciDoc is open and the FilterTec Plus is running, but no data is being collected, verify that you are using the correct cable and that it’s correctly installed. Also, ensure WinWedge is running by checking the System Tray for the WinWedge Icon; if it’s not there, click on the Setup Button and indicate the port you are using. If WinWedge can't access the Com Port, refer to the previous troubleshooting subject for help with Com Port errors.